Approximately three-quarters of patients who undergo THA are revised due to aseptic loosening caused by the release of submicron particulate debris from the bearing surfaces and subsequent host reactions within the effective joint space.3,4 A well-cemented component seals and minimizes the joint space, thereby delaying and minimizing the incidence of aseptic loosening while securing stable primary fixation at the time of index surgery.
Basics of Cementation
Polymethylmethacrylate (PMMA) is a self-curing polymer that polymerizes in an exothermic reaction. First used for orthopedic applications in 1958 by Sir John Charnley, PMMA functions as a gap filler and load distributor. PMMA is not a form of glue and does not have adhesive properties.5Tapered, polished stems show certain mechanical characteristics of cement based on physical time-dependant properties. The majority of cement mechanical studies were performed under laboratory conditions and room temperature, but THA is an in vivo condition where PMMA is subjected to the physiologic environment of body temperature, isotonicity, and cyclic loading. The glass transition temperature of PMMA is close to body temperature (373° K versus 310° K). In this glass transition temperature, PMMA behaves as a viscoelastic solid (Slide 1 and Slide 2) where creep and inverse creep share in decreasing stresses at the interfaces (Slide 3). In addition to acting as a load spreader, shock absorber, and load decoupler, PMMA also changes the nature of stress conditions in the cement-bone interface.6
Although PMMA has a high compressive strength, it has a weak shear strength. Maximum cement-bone interface shear strength is obtained by exposing and thoroughly cleaning strong trabecular bone and then forcing into it pressurized low viscosity bone cement.7
Medullary bleeding pressure at the proximal femur amounts to 36 cm H2O, which can create a blood film barrier between the bone and the cement (Slide 4). The bleeding pressure displaces Simplex P Bone Cement (Stryker Howmedica Osteonics, Allendale, NJ) away from bone up to 6 minutes from the start of mixing the cement. It is essential to keep the cement mixture under high pressure until it sets.8 The surgeon must know when the cement is set. Setting times vary according to the type of cement and the ambient temperature of the operating theater (Slide 5).
Low viscosity cement is desirable in the femoral canal because it is easier to use with injection devices. Low viscosity cement is also suitable for microinterlocking in the cancellous trabeculae of the proximal femur. Cement-bone interface pressure directly correlates with cement viscosity and is maximum distally, reaching 119 N/cm2 during stem insertion when a distal plug is used. If the distal plug is not accurately sized, then it migrates distally and leads to sudden, permanent loss of pressure (Slide 6).9
Tools of Cementation
A variety of tools were developed to ensure successful femoral cementation. Orthopedic surgeons have the following tools to use during cementation:
- Cement mixers
- Injection guns
- Distal femoral plugs
- Proximal seals
- Brushes
- Pulsed lavage systems
- Canal dryers
- Stem centralizers
Cement mixers decrease the porosity of cement and eliminate air bubbles. The mixers use centrifugal force, vacuum force, or a combination of the two. Cement mixers increase the cost of the procedure, the number of tools used during the procedure, and theater traffic. There is considerable clinical evidence that reduction of porosity in acrylic cement does not alter the long-term outcomes of THA (Slide 7).10
Injection guns are essential for retrograde installation of cement. They produce a continuous mass of cement fill in the proximal femur, and the barrel can be used in manual cement mixing (Slide 8).
Distal femoral plugs are used to close the chamber of the proximal femur. The plugs are made of polyethylene, PMMA, bioabsorbable gel, or bone plugs harvested from the femoral head (Slide 9).
Proximal seals are used after cement injection to keep the space of the proximal femur closed under pressure until the cement cannot be displaced by the medullary bleeding pressure (Slide 10).
Brushes, pulsed lavage, and gas-jet systems are used to remove the loose debris of cancellous bone and marrow in the proximal femur. Only bone trabeculae that are strongly fixed in the medullary cavity remain, allowing the maximum cement interlock with stable, living trabecular bed (Slide 11).7,11
Canal dryers are used to keep the distally plugged and clean canal dry of blood and marrow until the cement is mixed and injected. Canal dryers vary from ordinary packs soaked in thrombin or H2O2 to sponges or suction sieves (Slide 12).
Cement centralizers keep the stem in the middle of the canal, thereby eliminating varus or valgus orientation and maintaining a uniform cement mantle. The design adopted by the Exeter group ensures that the stem tip will be non-end bearing on the cement. Consequently, subsidence of the stem will not drive the tip into the cement, which could lead to wedging and fracture (Slide 13). Proximal centralizers are also used in other designs to keep an even mantle around the proximal femoral cut (Slide 14).
Pioneering work resulted in the evolution of techniques for femoral cementation through what is called cement generations. Although a standard is not set for each generation, the general layout is summarized in Table 2.12,13
Cementing Procedure
The steps of cementing the femur are best remembered in the following sequence:
- Planning
- Proximal cut & rasping
- Distal plugging
- Canal preparation
- Cement mixing
- Canal drying
- Retrograde cement injection
- Proximal sealing
- Stem insertion
- Setting
Planning must be considered. By using a template, a surgeon can establish the proper stem size that allows 2 mm to 5 mm of uniform cement mantle. The level of stem insertion is determined with the level of the neck cut (Slide 15). A template is also used to measure the canal diameter and determine the best femoral plug size for the patient.
Proximal cut is performed according to the template plan. This is important whether the stem is collared or not. Rasping with an appropriate serial-sized broach is done after a straight passage is created in the proximal femur by an initial broach introduced through the piriform fossa and prepared by a box chisel (Slide 16).
Distal plug insertion is performed over a graduated introducer, so it should be seated snuggly and fit about 20 mm distal to the proposed level of the distal stem tip (Slide 17).
Cement mixing starts with two packs of low viscosity cement. Time keeping is started at this point. The cement is ready for the assembled gun when it reaches a doughy state. Proper mixing ensures the removal of large air bubbles.
Canal drying occurs during the cement mixing by using one the devices mentioned (eg, a long cotton pack soaked in thrombin that is pressed and left in the canal until the gun is ready for injection).
Retrograde injection is started at the time specific to the cement type. The nozzle is inserted as deep as possible near the plug, and the cement is injected while the nozzle is withdrawn. As the nozzle comes out of the canal, it is broken on a previously etched mark just below the proximal seal. The proximal seal is pressed hard over the space of the proximal cut, and the rest of the cement is injected. Marrow fat and blood is forced out of the cancellous bone, a pointer to achieving a good pressurization level (Slide 10).
The proximal seal presses the cement in a closed chamber until it cannot be displaced by blood. Every cement brand has its own setting table that should be known by the surgeon.
The stem insertion consists of a surgeon placing the stem in the correct version with a steady motion while the thumb or a horseshoe-like seal maintains pressure over the calcar top. This pressure continues until cement setting is achieved (Slide 18). Before the full setting, any excess cement is removed. At the conclusion of the procedure, a pristine sealed interface is present between the stem neck and the femoral canal (Slide 19).
Conclusion
Training the operating staff on this stepped procedure ensures that reproducibility of a cemented femur occurs and that time is not wasted during the complex procedure. The quality of the cementation should be documented in the first postoperative radiogram, which improves the surgeon's learning curve and serves as a starting point for outcome analysis.14
References
- Harris WH. Options for primary femoral fixation in total hip arthroplasty. Cemented stems for all. Clin Orthop. 1997; 344:118-123.
- Smith SW, Estok DM, Harris WH. Total hip arthroplasty with use of second-generation cementing techniques. An eighteen-year-average follow-up study. J Bone Joint Surg Am. 1998; 80:1632.
- Malchau H, Herberts P, Eisler T, Garellick G, Soderman P. The Swedish total hip replacement register. J Bone Joint Surg Am. 2002; 84 (Suppl 2):2-20.
- Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: Polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg Am. 1992; 74:849-863.
- Timothy M, Wright SL. Biomaterials. In: Orthopaedic Basic Science: Biology and Biomechanics of the Muskuloskeletal System. Rosemont, Ill: AAOS; 2000:200-201.
- Lee AJC, Perkins RD, Ling RSM. Time-dependent properties of polymethylmethacrylate bone cement. In: Older MJO, ed. Implant Bone Interface. Springer-Verlag, London: 1990:85-90.
- Halawa M, Ling RSM, Lee AJC, Vangala SS. The shear strength of trabecular bone from the femur and some factors affecting shear strength of the cement-bone interface. Arch Orthop Trauma Surg. 1978; 92:19.
- Benjamin JB, Gie GA, Lee AJ, Ling RS, Volz RG.Cementing technique and the effects of bleeding. J Bone Joint Surg Br. 1987; 69:620-624.
- Oh I, Carlson CE, Tomford WW, Harris WH. Improved fixation of the femoral component after total hip replacement using a methacrylate intramedullary plug. J Bone Joint Surg Am. 1978; 60:608-613.
- Ling RS, Lee AJ. Porosity reduction in acrylic cement is clinically irrelevant. Clin Orthop. 1998; 355:249-253.
- DiGiovanni CW, Garvin KL, Pellicci PM. Femoral preparation in cemented total hip arthroplasty: Reaming or broaching? J Am Acad Orthop Surg. 1999; 7:349-357.
- Oh I, Harris WH. A cement fixation system for total hip arthroplasty. Clin Orthop. 1982; 164:221-229.
- Oh I, Bourne RB, Harris WH. The femoral cement compactor. An improvement in cementing technique in total hip replacement. J Bone Joint Surg Am. 1983; 65:1335-1338.
- Barrack RL, Mulroy RD Jr, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992; 74:385-389.