Hyperguides

Orthopaedics / Trauma

External Fixation

Gene Soeller, MD
Article

Introduction

External fixation has been used to treat fractures for more than 130 years. External fixation was first reported in 1853 by Malgaigne, who described using an external claw device to treat patellar fractures. In the 21st century, external fixation remains a valuable tool for providing temporary stabilization of acute fractures or dislocations, as well as treatment of nonunions or bony defects. In the management of severe musculoskeletal trauma, external fixators can also be useful for treating difficult periarticular fractures, particularly in the proximal and distal tibia, and for rapidly stabilizing pelvic and extremity fractures in multiple-injured patients.

Types of Fixators

Unilateral frame
When all pins in an external fixator are coplanar, the term used to describe the fixator is uniplanar (Slide 1). Uniplanar frames can be either unilateral or bilateral, depending on whether half pins or full pins are used.

Biplanar or delta frame
Biplanar frames have pins entering the bone typically at 90° angles to one another to provide increased stability (Slide 2).

Ring fixators
Ring fixators represent a modification of modular fixators and form a semicircular exoskeleton around the limb (Slide 3). Although ring fixators are the most cumbersome, most difficult to apply, and severely limit access to the leg for management of soft tissue injuries, they are the most versatile in terms of their function. They provide excellent mechanical stability and adjustability with the use of rods designed to allow gradual multiplanar correction of deformity. In addition, tensioned wires attached to a ring can provide secure fixation in even very small fragments of fractures.

Hybrid frame
The joining of a uniplanar or biplanar fixator with a ring to allow fixation of small metaphyseal or ephiphyseal fragments results in a hybrid frame.

Stability of Fixation

The stability of any external fixator is dependent on a number of factors including pin size, number of pins, pin distance from fracture site, configuration of frame, distance of bar from bone, bony contact, fixation in multiple planes, and double stacking.

Pin size
The bending strength of a pin is proportional to its radius to the fourth power. When pin diameters are too small, local stresses lead to micromotion and ultimate pin failure. When pin diameter is too large, the stress riser formed in the bone leads to fracture, even months after removal of the device. The increased rigidity achieved by using a larger pin is the most important way to increase frame rigidity.

Number of pins
Frame rigidity is increased by introducing additional pins and, more importantly, placing additional pins in a plane more than 45° from previous pins.

Pin distance from fracture site
The most stable frame is achieved when pins are placed as close to the fracture as possible (without entering the fracture site) and as far from the fracture site as possible.

Configuration of frame
A unilateral frame is the least rigid form of fixation. It resists forces best in the plane of the frame. Rigidity of this frame can be increased by double stacking the frame, or by placing another frame in a different plane (e.g., the delta configuration). Another way to increase frame rigidity is to use a circular frame.

Distance of bar from bone
The bar must be placed two finger breadths from the skin to allow for swelling. Placement farther from the skin decreases rigidity and allows additional micromotion. Over subcutaneous bones, such as the tibia, bars may be placed closer to the skin.

Bony contact
An important way to decrease the forces transmitted through the frame is to maintain bony contact at the fracture site. This allows some of the forces to be transmitted through the bone, which discourages bony resorption at the pin sites.

Bone Healing

Multiple factors influence the bone healing process including fixator rigidity, accuracy of restoration of length and alignment, fracture configuration, fracture loading conditions, and the degree to which the blood supply to the bone in the region of the fracture has been disrupted.

There are two types of fracture healing — union through external periosteal callus and primary bone healing, also known as in situ fracture remodeling. Fracture callus forms in reaction to the disruption of the periosteum and endosteum combined with the interfragmentary strain or motion associated with a bone injury. Callus bridges the fracture fragments and acts as a stabilizing structural framework and the biological substrate that provides the cellular material for union and remodeling.

The phenomenon of primary healing occurs in fractures that are rigidly stabilized. The rigid fixation precludes significant interfragmentary motion, thus necessitating remodeling by bony tunneling. Such injuries primarily remodel across contact points and minute fracture gaps. Remodeling, however, takes time. In fact, the rigid plates may not be safely removed from diaphyseal fractures until 12 to 18 months after fixation.

In an attempt to reproduce plate-like stability, early external fixation systems stressed the need for increasingly rigid frames in multiplanar configurations. These techniques may have actually delayed or prevented union by preventing the formation of bridging callus. A rigid external fixator that eliminates micromotion must be kept in place longer and requires prolonged maintenance of the fixator pin/bone interface. When an external fixator is applied, a "race" begins between fracture healing and fixator pin failure (due to infection, loosening, etc). Current external fixation systems are designed to allow micromotion at the fracture site to promote callus formation. Stable yet less rigid systems of external fixation maintain alignment and length while allowing beneficial micromotion.

Pin Insertion

The method of pin insertion is one of the most important determinants of long-term stability of the pin. The insertion technique may have mechanical effects on the initial screw/pin purchase as well as a profound biological influence on the maintenance of the bone-implant interface. The insertion of self-drilling screws is associated with microfracture and high temperatures at the bone-implant interface. Thermal necrosis often results. The tensioned K-wires used with circular-ring external fixation techniques are prone to thermal necrosis when inserted in hard diaphyseal bone. To reduce the potential for both thermal necrosis and the premature loosening associated with it, an alternative technique that involves predrilling pilot holes for all screw or pin sites is now recommended to minimize microfracture and avoid excessive bone temperatures.

Preloading pins by bending them was suggested as a means of augmenting the tightness of the pin within the bone. In a study comparing the effects of bending preload with radial preload in reducing resorption of bone at the pin-bone interface, researchers found that radial preloading was superior to bending preloading. High degrees of radial preload that exceed the elastic limit of cortical bone may result in microfracture and subsequent resorption. Accurate predrilling of holes before pin insertion is essential. Holes should be predrilled at low speed with a fresh, sharp drill point. Pins must then be inserted by hand, but take care and do not allow the pin to toggle during insertion because this may cause microfracture and decrease pin torque resistance. If pins with threads that engage both the near and the far cortices of the bone are chosen, then the drill diameter should match the shank diameter of the pin. If pins with short threads that engage only the far cortex of the bone are selected, then the drill diameter should be 0.2 mm less than the shank diameter of the pin to provide for adequate radial preload at the near cortex without causing microfracture of the surrounding cortical bone.

Complications

To accurately appreciate the magnitude of pin tract infections, infections must be graded or staged. Each stage characterizes a point in the continuum of the pathophysiological "natural history" of colonization, bacterial overgrowth, early soft tissue infection, later established deep infection, and osteomyelitis.

Stage I: Inordinate serous or seropurulent drainage
All pins will drain, but normal drainage must be as minimal. Free-flowing drainage is pathologic, as is any degree of purulence. These infections require treatment with aggressive pin-site hygiene and oral broad-spectrum antibiotics, usually involving a first-generation cephalosporin.

Stage II: Superficial cellulitis
A halo of cutaneous erythema that extends from the pin site usually indicates a soft tissue infection. At this stage, the infection requires oral antibiotics and increased pin-site hygiene. If this treatment meets with a discouraging response, parenteral antibiotic treatment may be indicated.

Stage III: Deep infection
This stage is uncommon and is characterized by deep-seated infection along the entire pin track. It can be differentiated from superficial pin-site cellulitis by purulent drainage, swelling, and severe cellulitis encompassing more than one of the pins in a cluster. Antibiotic treatment is required, often via the parenteral route. Such infections raise the suspicion of bone infection and implant loosening. Loose implants must be removed. Pins or screws may need to be replaced. Reinsertion must only be performed after treatment with a course of antibiotics to "sterilize" the soft tissue sites chosen for new pin implantation. Introducing another implant through or near acutely cellulitic tissue invites another infection by inoculating the soft tissue and bone of the new site with the infecting organism.

Stage IV: Osteomyelitis
Clinical loosening of implants accompanied by infection and radiographic evidence of bone involvement must be interpreted as infection of bone. Acute infection is often treated by hardware removal and 10 to 14 days of parenteral antibiotics. If radiographic evidence of a ring sequestrum exists with persistent or recurrent drainage despite antibiotic treatment, surgical debridement is performed by overdrilling the infected pin sites with a drill bit larger than the screw diameter on a hand brace while providing copious pulsed suction/lavage and curettage of the defect. Bone specimens, not swabs of drainage, should be sent for culture. Postoperative oral antibiotic treatment for 14 days is sufficient after the pin track is adequately debrided.

Pin loosening
Gradual mechanical pin loosening is, to some extent, a natural process in external fixation because pins or screws must withstand chronic stress. Loosening can be minimized with proper insertion techniques that stress predrilling of screw holes, radial preload (avoiding bending preload), euthermic pin insertion, and adequate soft tissue release around the implant sites. All implants eventually loosen if subjected to continued stress.

Malunion
The outcome of external fixation is only as good as the initial alignment of the fracture fragments. Every attempt must be made to ensure a near anatomic alignment upon placement of a frame. Close follow-up is necessary to ensure that uniplanar frames do not drift due to application of physiologic forces.

Nonunion
Delayed union and nonunion were once commonly cited as likely outcomes associated with the use of external fixation. However, by better understanding the biology of fracture healing and the potential causes of nonunion, clinicians have achieved union rates with external fixation comparable to other internal fixation techniques for injuries of like severity. Close attention to restoration of bone length and bone grafting of defects helps to ensure timely union.

Bibliography

Behrens F, Searls K. External fixation of the tibia. Basic concepts and prospective evaluation. J Bone Joint Surg Br. 1986; 68:246.

Green S. Complications of pin and wire external fixation. Instr Course Lect. 1990; 39:219-228.

Halsey D, Fleming B, Pope MH, et al. External fixator pin design. Clin Orthop. 1992; 278:305.

Hydahl C, Pearson S, Perren SM. Induction and prevention of pin loosening in external fixation: An in vivo study on sheep tibiae. J Ortho Trauma. 1991; 5:485.

Jupiter JB, Levine AM, eds. Skeletal Trauma. Philadelphia, Pa: WB Saunders; 1992:110-112.

McBroom RJ, Cheal EJ, Hayes WC. Strength reductions from metastatic cortical defects in long bones. J Orthop Res. 1988; 6:369.

Nepola JV. External fixation. In: Rockwood CA Jr, Bucholz RW, Green DP, eds. Fractures in Adults. 4th ed. Philadelphia, Pa: Lippincott-Raven; 1996:229-259.

Paley D. Problems, obstacles, and complications of limb lengthening by the Ilizarov technique. Clin Orth. 1990; 250:81.