The posterior cruciate ligament (PCL) is saved approximately 60% of the time and substituted 40% of the time. When substituting, a more conforming, deep-dish polyethylene can be used.
If the PCL is going to be saved, it must be protected during proximal tibial resection. This can be accomplished by placing a ¼-inch osteotome in front of the PCL prior to the saw cut. If this is done, the PCL can be saved 100% of the time. To preserve the PCL, exact tensioning of the knee is critical. This can be accomplished by a measured resection technique of the femur, tibia, and patella, and reproducing the normal posterior slope of the tibia. Using this surgical technique, the PCL will have good tension throughout the range of motion.
Two choices are available for substitution: a post or deep-dish polyethylene. Complications reported with the use of posts include patellar fractures, knee dislocations, and patellar clunk syndrome. Raising the joint line can cause mid-flexion instability, which can lead to dislocation.
Patellar clunk syndrome can be avoided by using a deep-dish insert. Patellar clunk occurs from a larger intercondylar notch of the prosthesis, which causes fibrous tissue build-up proximally and can result in 1% to 2% of patients requiring arthroscopic debridement.
Investigational work (unpublished data, A.S. Greenwald) has evaluated the stability of a deep-dish polyethylene insert to substitute for the PCL. In both flexion and extension, this device with a 12.5-mm anterior build-up provided excellent stability and greater strength compared to the PCL during laboratory testing. Standard inserts were inferior and did not effectively substitute for the PCL.
Hofmann et al1 reviewed their use of ultracongruent polyethylene over 7 years in 100 patients who underwent PCL-substituting total knee arthroplasties. Fifty-three cases were primary and 47 were revisions. There were no cases of anteroposterior (AP) instability in either revision or primary cases when a deep-dish polyethylene was inserted. The incidence of AP instability using standard inserts was 2% to 3%.
The decision to use a deep-dish insert is made intraoperatively. After placing the trials, the PCL is tested. If there is evidence of AP instability, the ultracongruent insert is used as a substitute.
The ultracongruent insert provides excellent stability when performing a forced posterior drawer test. Stability is present at more than 75° (e.g., the standard stability that occurs with a standard posterior stabilized insert), as well as at 0°, 15°, and over the entire range of motion.
The ultracongruent insert provides excellent posterior stability and has fewer complications than the traditional posterior stabilized. It is technically easy to insert and is bone spared, making it a logical choice.
References
- Hofmann AA, Tkach TK, Evanich CJ, Camargo MP. Posterior stabilization in total knee arthroplasty with use of an ultracongruent polyethylene insert. J Arthroplasty. 2000; 15:576-583.