Haim Stein, MD, D. Phil (Oxon) Stephan M. Perren, MD, DSc (HC) R. Mosheiff, MD F. Baumgart J. Cordey, DSc Reprinted with permission from Stein H, Perren SM, Mosheiff R, Baumgarts F, Cordey J. Decline in transfixing k-wire tension of the circular external fixator: Experimental continuous in vivo measurements. Orthopedics. 2001; 24:985-989.
Article

The mechanical properties of an external fixation frame, used to treat any bone defect, determine the biomechanical environment in the healing bone gap.1-11 Mechanical properties of various external fixators have been studied in vitro4,5,9,12-17 and the magnitude of their stiffness may affect the outcome of treatment.18-20 Dynamization, a clinical term used to describe induced micromovements between opposing bone fragments, seems to enhance bone healing.21 However, no precise or reliable method has yet been described for measuring or controlling such interfragmentary in vivo movement.22,23

The high success rate of the Ilizarov system in inducing rapid and effective bone healing is attributed to its unique mechanical properties, which support effective osteoneogenesis in distraction.6 The main difference between the Ilizarov system and uniplanar external fixators is in the latter, bone is transfixed by cantilever pins that articulate with a uniplanar frame. In the former, thin Kirschner wires transfix the bone and need to be stabilized in a three-plane configuration. When the bone is axially loaded, these thin K-wires show a nonlinear relationship with frame stiffness under varying degrees of load, hence, the K-wires act as "strings" not bending beams.14-16,24 The tension in the transfixing K-wires has been distinctly identified as a crucial factor in determining frame stiffness.8,23,25

This article reports alterations measured continuously in such tensions in vivo. The experimental procedure was comparable to similar clinical conditions. The results of this study provide new data on the crucial aspect of distraction osteoneogenesis.

Materials and Methods

To define the material properties of the K-wires, they were individually studied. Measurements were performed on 10, 1.25-, 1.6-, and 2-mm diameter AO/ASIF K-wires. Two wires of each size were investigated in a mechanical testing machine (Instron, Canton, Mass) in the traction mode. The wires were fixed between two clamps; the segment of wire between the clamps measured 10 cm.

A displacement rate of 10 mm/minute resulted in change of the applied tension force. Once the predetermined tension was achieved, the wire was secured by tightening the nut connected to the load cell (Slides 1 and 2).

In clinical and experimental circumstances, where no load cells are used, the wires are tensioned tightly around the screw head after tightening the nut. This was the original Ilizarov procedure. Whether any tension was lost during this phase was unknown and it was not documented in the literature. Because the in vitro results consistently showed a marked drop in tension over time, alterations in K-wire tensions were studied in vivo in experimental animals.

Two adult female Swiss mountain sheep, aged 5-7 years, weighing 59-62 kg, were operated on under general anesthesia with endotracheal intubation and controlled respiration. A modified circular four-ring external fixation frame was mounted on the left tibia, using 1.6-mm diameter stainless steel K-wires (Synthes #29218; Synthes, Paoli, Pa).

Percutaneously, the bone was predrilled with a 1.5-mm drill and the K-wires were then introduced, fixed at one end to the ring with a specially designed 6-mm screw, which had a fitting hole, drilled transversely to the longitudinal axis 1 mm below the screw head. A riled washer was placed just above the ring, with the riled surface facing the wire to optimize the grip similar to accepted industrial techniques. The hold on the wire was achieved by a nut tightened on the screw end protruding underneath the ring. Following the tightening of the nut, the protruding end of the wire was turned around the screw head with a pair of pliers.

On the opposite end of the ring, the wire was then threaded through a similar screw, tensioned with a special tension device (Slides 1), and the tension was recorded on an analog X-Y chart recorder (Graphtec Co Ltd, Yokohama, Japan). Special attention was paid to the issue of calibrating the entire calibrating chain. To allow continuous recording of this tension, a special instrument load cell was used to replace the screw head ( Slide 2). When the desired tension was reached, the nut was tightened.

One load cell was placed on each K-wire, two wires in each ring, four rings in each fixation frame. Signal monitoring was performed using a bridge amplifier. The tibia was positioned in the center of each ring, with the wires crossing at right angles.

Following frame stabilization, the tibial diaphysis was exposed through a longitudinal incision on the anteromedial border where bone is covered only by skin and subcutis. The periosteum was incised longitudinally and retracted and the bone was exposed. With a standard maxillofacial oscillating saw (Synthes AG), a segment of 2.5-cm bone was resected through two parallel osteotomies. Blood loss was minimal and was stopped by applying superficial pressure. The bone ends were brought into contact by acute shortening, and the wound was closed in layers by interrupted sutures.

Anesthesia was terminated, and when the sheep regained full consciousness it was returned to its enclosure where it was supported in a hammock to steady it and prevent sudden peak loading, which occurs in animals changing from a recumbent to an upright position. Similar peak load alterations are not expected in humans.

All K-wires were tensioned to 500 N, which is in the linear range of the load cell's calibration curve. The mechanical distraction was performed every six hours after a resting period of 72 hours at the beginning of the distraction. Distraction of experimental bone grafts was started at a rate of 0.25-mm six hourly for 28 days. At the end of that time, the original anatomic length of the operated tibia had been restored and the sheep returned to symmetrical full weight bearing. The external fixation frame was left in place for an additional 7 days for bone consolidation.

On postoperative days 14, 28, and 35, control radiographs were taken of the operated tibia at 55 Kv and 200 mA with an exposure time of 0.25 seconds. On postoperative day 35, following euthanasia, the left tibia was explanted and underwent morphologic studies.

Results

Continuous tension measurements were available for assessment in seven of eight wires in each sheep. In the eighth wire, the connecting cable wires broke in both sheep and could not be repaired for technical reasons. In both sheep, only measurements of the most distal ring were affected.

Within 2 weeks postoperatively, tension had declined significantly, reaching 0% in two wires and less than 20% of the original value in a third wire. At the end of 5 weeks, the tension in five of seven wires had declined to zero (Slide 3). This decline in tension occurred in all wires, but those in ring two from the top appeared to be most affected in both sheep. The graphic display of the forces interacting in this system is shown in Slide 4 and delineated in the insert. Under continuous axial loading, these loading forces induce a decline in the tension of the transverse-positioned K-wires (i.e., this interaction explains why the K-wire tension declines spontaneously).

Sequential radiographs showed the first signs of intramembranous ossification on postoperative day 14 (Slide 5). The surgical gap was filled with woven bone by postoperative day 35. On macroradiographies (500-µm thick sections), significant resorption was seen in cortical bone surrounding wire holes (Slide 6).

The mechanical properties of the individual K-wires are demonstrated in Slide 7. For the 1.6-mm wire, the elastic limit of the applied tension is reached at approximately 2000 N, whereas the failure limit is reached at approximately 3000 N. This means that the tensions applied to these wires while mounted on the sheep were well below their plastic deformation in this mode (i.e., they were within the elastic range).

Discussion

The three-plane configuration of the circular external fixation frame determines the displacement between the fracture fragments during weight bearing. All or most of the load is transmitted through the fixator.4

In clinical practice, selecting the frame configuration with the optimal stiffness means inducing the optimal displacement at the fracture site that will enhance healing.4 The inter-relationship between fixation frame stiffness and the developing callus determines the load share transmitted through each to the limb on weight bearing. The stiffer the callus, the higher the percent of load transmitted through it on loading.

When axial loads are applied to wires tensioned in a ring fixation, the initial wire tension first increases, and on removal of the load is reduced in relationship to the magnitude of this load, independent of the method of clamping used to fix the wire to the ring.5 In this study, continuous measurements in vivo of persisting tension in the transfixing wires were assessed for the first time. The findings were identical in both sheep; therefore, statistics are meaningless in this case.

These wires were fixed to the ring by a screw/nut locking mechanism. Under repeated high axial load, any tensioned wire positioned in a transverse plane to this load will either slip or undergo plastic deformation. This has been reported4,5,23 and also is shown by the present results. Therefore, the stiffness of the ring external frame, which is closely related to K-wire tension, is reduced by weight bearing over time. This final decrease in wire tension depends on the degree of plastic deformation induced by repeated axial loading. Other variables that affect frame stiffness when treating a fracture are the ring thickness and diameter, centric or eccentric location of the bone in the ring frame, and length and elasticity of the transfixed bone segment.

The technique of predrilling bone prevents extensive thermal cortical bone damage and subsequent loosening of the transfixing device.17,20,22,26,27 Load transfer between wire and bone depends mostly on the mechanical tension in the former, provided it is positioned at 90° to the bone axis. Any deviation from this angle will cause further loss in wire tension with any amount of fragmentation, biological bone resorption, or both around the wire developing in the aftermath of drilling and the heat generated by it.17,20,26 Heat generated when drilling cortical bone is due to friction and the fragmentation of bone particles.20 Although cortical circulation is effective,27 predrilling is crucial in limiting excessive temperature elevations.20 Such elevations in temperature have been shown to cause bone necrosis.28 A smooth pin or K-wire has no provision to eliminate debris accumulating between it and the drill-hole wall,20 whereas low drilling speeds cause an increase in fragmentation.27 Both conditions are met by the surgical techniques applied when mounting the Ilizarov frame. Therefore, it is likely these mechanical and thermal effects are further contributing factors to the encountered spontaneous decline in K-wire tension.

This study supports previous reports that reduction in frame stiffness is an important mechanical stimulus in promoting new bone formation.2,7 Despite this loss in tension, new bone formation appeared in the bone gap at 14 days postoperatively and was abundant at the end of 35 days.

On the basis of experimental studies, it has been proposed that strain applied to bone is a biological transmitter that "informs" skeletal tissues of mechanical demands6,10; therefore, bone mass is regulated by mechanical strain.9 Intermittent load has been found to be a stimulating factor for periosteal new bone formation.8 In clinical practice, this is achieved by using the Ilizarov system, which has a proven high rate of early callus formation and a nonlinear relationship between frame stiffness and K-wire tension.7,15,16,23 The results of the present in vivo study demonstrate that the tension in these wires begins to decline within 2 weeks of application and reaches zero levels by 5 weeks, although their material properties are capable to withstand much higher loads. Kirschner-wire fixation to the external frame, although performed by a screw/riled washer/nut gripping system, appears to slip with time under axial loading. This also contributes to the observed decline in wire tension under these conditions. Thus, it is shown that during the early phase of bone healing, the thin wire ring frame gradually loses stiffness due to many causes, producing a mechanical environment that seems to effectively stimulate the early phase of bone healing.2,7,21

Frame stiffness values provide only indirect information about the micromovement that occurs at the bone gap.10 These micromovements are crucial in promoting the bone healing process. Accurate data for the evaluation and assessment of the actual degrees of micromovement in vivo are not available. These micromovements enhance the development of ample mineralized callus and woven bone, and the stiffness of the latter will replace the mechanical stiffness of the fixation frame, which spontaneously declines.

A cause of biological failure in bone healing often is inadequate mechanical stability of fractured bone ends in the early weeks of osteoneogenesis.24 The ring fixator, with its mechanical properties, creates a mechanical environment that supports and enhances the biology of bone regeneration during that critical and sensitive early phase. Effective callus neogenesis produces the ideal conditions of increased stability and stiffness of the proliferating woven bone mass, which effectively replaces the declining stiffness of the fixation frame. This was evaluated in theoretical and experimental studies.4,5,7,23,29,30-33 In the late stage of healing, the regenerating tissues are stiffer due to their regained biological integrity. The biomechanical properties of bone regenerate are unknown.29 The external ring fixation frame has the potential to develop both a high and low mechanical stiffness subject to the tension present in the K-wires at that time.

What is the tension required to elicit the optimal biological response? Ilizarov8 and Bagnoly et al34 recommended 80-90 kg, which correspond to 800-900 N accordingly, whereas Cattaneo et al35 used 130 kg in full rings and 90 kg in half rings. Our data show the monitored tensions significantly declined within 21 days postoperatively and after 35 days reached zero in six of seven measured wires. It would therefore appear that the original K-wire tension should be "optimal" and this "optimal" range appears to be at 1000 N.23,28

The measured decline in wire tension is a multifactorial event due to wire slippage, plastic deformation of the wire, and the effect of directly drilling bone with K-wires. These findings support the definition of Burny et al36 of the "elastic fixator," but although it provides an adequate explanation for the early phase of bone healing, it does not provide a solution for the later phase, i.e., the phase of bone remodeling.

In the literature, the reported data show no difference between various external fixation systems with regard to the overall length of treating time needed for complete fracture healing (i.e., to the end of the remodeling phase).1,13,14,37-41 Our observations lend support to the conclusion of Wu et al31 that the search for the ideal fixator is still open, but according to current knowledge, the hybrid ring tubular frame seems to meet many of these prerequisites.23

References

  1. Aro HT, Kelly PK, Lewallen DG, Chao EY. Comparison of the effect of dynamization and constant rigid fixation on rate and quality of bone osteotomy union in external fixation. Transactions of the 34th annual meeting of the American Orthopaedic Research Society; 1988:303 (abstract).


  2. Aronson J, Harp JH Jr. Mechanical considerations in using tensioned wires in a transosseous external fixation system. Clin Orthop. 1992; 280:23-29.


  3. Cheal EJ, Mansmann KA, DiGioia AM III, Hayes WC, Perren SM. Role of interfragmentary strain in fracture healing: ovine model of a healing osteotomy. J Orthop Res. 1991; 9:131-142.


  4. Delprete C, Gola MM. Mechanical performance of external fixators with wires for the treatment of bone fractures, I: Load-displacement behavior. J Biomech Eng. 1993; 115:29-36.


  5. Delprete C, Gola MM. Mechanical performance of external fixators with wires for the treatment of bone fractures, II: Wire tension and slippage. J Biomech Eng. 1993; 115:37-41.


  6. Goodship AE, Kenwright J. The influence of induced micromovement upon the healing of experimental tibial fractures. J Bone Joint Surg Br. 1985; 67:650-655.


  7. Goodship AE, Watkins PE, Rigby HS, Kenwright J. The role of fixator frame stiffness in the control of fracture healing. An experimental study. J Biomech Eng. 1993; 26:1027-1035.


  8. Ilizarov G. Clinical application of the tension-stress effect for limb lengthening. Clin Orthop. 1990; 250:8-26.


  9. Podolsky A, Chao EY. Mechanical performance of Ilizarov circular external fixators in comparison with other external fixators. Clin Orthop. 1993; 293:61-70.


  10. Rubin CT, Lanyon LE. Regulation of bone mass by mechanical strain magnitude. Calcif Tissue Int. 1985; 37:411-417.


  11. Williams MO. Long-term cost comparison of major limb salvage using the Ilizarov method versus amputation. Clin Orthop. 1994; 301:156-158.


  12. Calhoun JH, Li F, Bauford WL, Lehman T, Ledbetter BR, Lowery R. Rigidity of half-pins for the Ilizarov external fixator. Bull Hosp Jt Dis. 1992; 52:21-26.


  13. Fleming B, Paley D, Kristiansen T, Pope M. A biomechanical analysis of the llizarov external fixator. Clin Orthop. 1989; 241:95-105.


  14. Fischgrund J, Paley D, Suter C. Variables affecting time to bone healing during limb lengthening. Clin Orthop. 1994; 301:31-37.


  15. Gasser B, Wyder D, Schneider E. Comparative investigation on the biomechanical properties of the circular and other three-dimensional fixators in biomechanics. In: Bergmann G, Kolbel R, Rohlmann A, eds. Basic and Applied Research. Netherlands: Martinus Nijoff Publishers; 1987:429-434.


  16. Gasser B, Boman B, Wyder D, Schneider E. Stiffness characteristics of the circular Ilizarov device as opposed to conventional external fixators. J Biomech Eng. 1990; 112:15-21.


  17. Lipkin M, Hardy J. Measurements of some thermal properties of human tissues. Appl Physiol. 1954; 7:212-217.


  18. Karaharju EO, Aalto K. The deformation of external fixation devices during loading. Int Orthop. 1983; 7:179-183.


  19. Kempson GE, Campbell D. The comparative stiffness of external fixation frames. Injury. 1981; 12:297-304.


  20. Matthews LS, Hirsch C. Temperatures measured in human cortical bone when drilling. J Bone Joint Surg Am. 1972; 54:297-308.


  21. Lanyon LE, Rubin CT. Static vs dynamic loads as an influence on bone remodeling. J Biomech. 1984; 17:897-905.


  22. Schmelzeisen H. Mechanische, Thermometrische und Biologische Untersuchungen beim bohren in der Corticalis von Rohrenknochen [habilitations these]. Freiburg, Germany: University of Freiburg; 1988.


  23. Stein H, Mosheiff R, Baumgart F, Frigg R, Perren SM, Cordey J. The hybrid ring tubular external fixator: A biomechanical study. Clin Biomech (Bristol, Avon). 1997; 12:259-266.