Introduction
Spinal fusion remains an important treatment option for numerous pathologies. However, achieving a successful fusion is challenging, and rates of pseudarthrosis have been reported as high as 60%.1 A number ofstudies have linked solid fusion to improved clinical outcomes.2,3 Various treatment adjuncts have been developed to increase fusion rates and reduce the incidence of pseudarthrosis. Treatmentoptions include spinal instrumentation, bracing, osteoinductive factors, and electromagnetic stimulation (EMS).
Electromagnetic stimulation has been used to facilitate long bone healing for some time. Recently, a growing body of literature has begun supporting EMS as an adjunct to spinal fusion, especially in cases where low fusion rates are anticipated. A significant number of variables are poorly defined and a lack of guidelines for EMS use has hampered widespread use in spinal fusion.
Background
Although the use of EMS in bone healing was first reported in 1841, it was not until the 1970s that serious interest began to develop in this modality.4 Most of the early research focused on the treatment of long bone nonunions, and the U.S. Food and Drug Administration (FDA) first approved electrical stimulation devices for this application in 1979. In 1974, the first clinical use of electrical stimulation in spinal fusion was reported,5 and the FDA first granted approval of this technology for spinal application in 1987.
Electric Physiology of Bone
Bone tissue has an electrical potential known as the bioelectric or steady state potential, which is dependent on the bone tissue’s viability.6 Bone cells generate this potential by active pumping of electrolytes.7 This potential is negative in areas of bone growth or healing and will gradually return to a neutral or electropositive state as healing progresses and abates.8
Bone also exhibits a different type of electric potential depending on the forces applied. Compression tends to result in electronegative potentials and tension tends to result in electropositive potentials.9-11 One theory suggests that bone tissue interprets stress through electrical signals and that compression and tension potentials form feedback signals that help modulate growth and remodeling, allowing bone to adapt to different mechanical stresses.12
Three modalities are used to electrically stimulate bone:
- Direct current electrical stimulation (DCES)
- Pulsed electric magnetic field (PEMF)
- Capacitively coupled electric stimulation (CCES)
Direct Current Electrical Stimulation
Direct current electrical stimulation works by delivering an electric current (usually in the range of 20 milliamperes to 40 milliamperes) directly to the site of nonunion or the fusion bed through two titanium electrodes implanted into the fusion bed. These electrodes function as the cathode (the negative pole). The electrodes connect to a battery pack, which is implanted in a pocket of tissue in the paramedian region no more than 8 cm to 10 cm away and functions as the anode.
The effective area of stimulation is 5 mm to 8 mm from the electrode. The implant can be coiled to increase the surface area.12The electrodes are generally placed in the fusion bed after decortications and before transfer of bone graft. If instrumentation is being used, then care is taken to separate the electrodes from the hardware with soft tissue. Resultant bone growth shows a dose-dependant repose when stimulated with currents between 5 microamperes and 20 microamperes. Lower currents are insufficient to induce growth, whereas higher currents increase the risk of necrosis.13,14 Recently, devices have been tested using higher microamperes in animals.1,15
Stimulation is generally used from 6 to 9 months, and the battery pack can be removed afterward (but is often left in place). The long-term safety of retaining the battery pack and electrodes was confirmed in one study over a 10-year period.16 Two studies have discussed the safety of DCES devices during magnetic resonance imaging (MRI), and at least one supplier of DCES has changed its guidelines to allow MRI of patients with these devices.17,18 Nevertheless, it is generally recommended to remove the battery prior to MRI studies due to concern of interaction.
Mechanism of action
Electromagnetic stimulation has an effect on the growth of collagenous tissues; however,
the exact mechanism of action remains unclear.12,19,20 Some papers have suggested
different mechanism of actions, but no one theory predominates suggesting that more than one
mechanism of action might be at play.
With DCES, in vivo studies have demonstrated induced osteogenesis and stimulated bone healing at the negative (active) electrode if implanted at the site of the fracture.12,14, 21,22 In vitro studies have shown stimulation of mitosis and recruitment of osteogenic cells in cultures.23
Implanting an electrode results in physiological consequences:
- Attracting proteins and growth factors (electrophoresis)
- Migration of bone, cartilage, and endothelial cells to the fusion site (galvanotaxis)
- Polarization of cell membranes (voltage-gated channels and activation of cAMP cascade)
Different theories with varying bodies of supporting literature have been suggested. Oxygen depletion, stimulation of cAMP, and micromotion have been cited as potential mechanisms. The oxygen depletion theory is based on the fact that one of the reactions occurring at the active electrode (the cathode) results in oxygen consumption and an alkaline pH, which would favor bone formation. The cAMP theory suggests that electric stimulation directly increases the level of cAMP, activating the cascade of enzymes associated with cAMP culminating in osteogenesis.12,24 The micromotion theory suggests that the electrode undergoes small movements that stimulate osteogenesis and not the electric current itself.25
Experimental studies
Direct current electrical stimulation efficacy was studied initially in animal models,
including swine and dog, and was effective.5,15,26-28 As far as clinical trials,
Kane was the first to study DCES in spinal fusion with a prospective, historically matched
study in 1981 followed by a patient-blinded, randomized, controlled trial in 1988.27
His latter study looked at the effect of DCES in what was defined as "difficult
patients":
- Patients undergoing revision fusions
- Grade II or worse spondylolistheses
- Multiple-level fusions
- Presence of another high-risk factor, such as obesity
In this study, 59 patients were followed for 12 to 18 months postoperatively. Fusion was determined radiographically as assessed by the surgeon and an independent radiologist. Eighty-one percent of patients with DCES devices went on to fuse compared to 54% in patients with placebo. This study concluded that DCES promoted a fusion (a statistically significant difference).
The other randomized controlled trial was performed by Rogozinski and colleagues.28 They looked at the effect of DCES on patients undergoing instrumented lumbar fusions for low back pain. Patients were followed for an average of 20.5 months. Fusion was determined radiographically as assessed by the orthopedist. Ninety-six percent of the patients with DCES devices went on to fuse compared to 85% in patients with only instrumentation. This study concluded that DCES improves fusion rates (statistically significantly). This study had a number of weaknesses as only 26 of its 94 patients were randomized, and fusion was assessed by the treating orthopedist with no independent judge.
There have also been a number of retrospective clinical studies on DCES. In these studies, fusion rates for groups with DCES ranged from 93% (vs. 75% in historical control, P=.0003) in Meril’s study29 to 95.6% (vs. 87% in historical control, P=.05) in Kuchrzyk’s study.30 Overall, these studies suggest that DCES increases rates of spinal fusion methodology.
Pulsed Electric Magnetic Field (PEMF)
A pulsed electric magnetic field works by delivering an alternating electric current through an external coil, thus generating an electromagnetic field that is constantly changing in a defined manner. The preferred pattern is known as a bipolar, quasirectangular base pulse pattern - a pattern that is asymmetric in time and amplitude. The coils are usually incorporated into a brace that the patient wears for 3 to 8 hours a day for 3 to 6 months. The placement of the coils in relation to each other is crucial and should center on the area of desired effect. Although the effect of instrumentation was initially a concern, it does not seem to have a dampening effect of PEMF efficacy provided that the braces does not shield the fusion bed.12,31
A sub-variant of PEMF is a combined magnetic field (CMF), which differs from PEMF in that it superimposes the magnetic field's pattern of change on a pre-existing static magnetic field. Combined magnetic field devices are worn for 30 minutes a day. This number comes from an animal study that showed increased bone stiffness at 30-minute doses, yet also showed a much larger response when a 24-hour per day dose was used. A comparison of CMF with PEMF in a rabbit tibial osteotomy model showed that both have the same efficacy when used for the same duration.15
Mechanism of action
Researchers hypothesize that the biological activity of PEMF is due to the electric current it
generates or induces with tissue stimulation. Among the postulated mechanisms are: changes to
cell membrane potentials or changes to the molecular configuration of the parathyroid hormone
receptor, increased calcium influx into bone cells,15,32-34 and increased levels of BMP-2 and
BMP-4 mRNA.35,36
Experimental studies
Studies of PEMF efficacy in animals were initially negative because they failed to show any
statistically significant increase in fusion rates in dog models. However, PEMF efficacy was
shown in a rabbit model and in human clinical trials. Various studies have looked at the
efficacy of PEMF in spinal fusion. Of these studies, we mention two randomized controlled
trials.
Mooney's double-blind, randomized study evaluated 195 patients undergoing anterior or posterior interbody spinal fusion into two groups. One group wore a brace with a PEMF device, and the other group wore a sham brace. Patients were instructed to wear the brace for at least 8 hours a day. Patients were followed for a minimum of 12 months and were assed radiographically by the orthopedist and an independent radiologist, as well as clinically by the orthopedist. In the PEMF group, 92% of patients fused compared with 65% in the control (P=.005). The study concluded that PEMF is effective for bone graft stimulation.37
Linovitz and colleagues38 also conducted a double-blind, randomized study that included 201 patients undergoing spinal fusion. Patients were placed into one of two groups. One group wore a PEMF device, and the other group wore a sham device for 30 minutes a day for 9 months. The endpoint was 9 months after surgery. Fusion was assessed radiographically as judged by a blind panel consisting of the surgeon, a musculoskeletal radiologist, and a spine surgeon. Fusion rate in the PEMF group was 67% compared to 43% in the control group (P=.003). The study concluded that PEMF devices significantly increase the success of spinal fusion.
Capacitively Coupled Electric Stimulation (CCES)
Capacitively coupled electric stimulation works by applying an electric potential to two external plates separated by a distance, forming a magnetic field. Whereas PEMF devices apply a magnetic field that is changing in intensity, CCES devices apply a field that is constant. The device usually consists of a small computer connected through electrodes on either side of the spine at the level of the fusion. Capacitively coupled electric stimulation devices are usually worn constantly for 9 months or until fusion is achieved. The electrodes are replaced periodically.12
Mechanism of action
A number of postulated theories explain the mechanism of action of CCES. One theory suggests
that CCES causes the movement of calcium across cellular membranes through voltage-gated calcium
channels, resulting in the activation of calmodulin and increased levels of prostaglandin E.
Other evidence points to increased levels of transforming growth factor beta 1 after
CCES.39-41
Experimental studies
To our knowledge, there has only been one randomized controlled trial examining CCES efficacy on
spinal fusion. This was a double-blind study of 179 patients conducted by Goodwin and
colleagues.42 One group of patients wore the CCES device, and the other group of patients
wore a sham device. The patients were instructed to use the device continuously. Through a
special feature on the device, the investigators were able to calculate the actual average
daily usage of the device, which was 15.7 hours a day for the control group and 16.5 hours for
the placebo group. Patients were evaluated radiographically by the orthopedist and an
independent radiologist, as well as clinically by the orthopedist. The fusion rate in the group
wearing CCES devices was 84.7% compared to 64.9% in the placebo group (P=.0043). The study
concluded that CCES is an effective adjunct to primary spine fusion, especially for patients
with posterolateral fusion and internal fixation.
Discussion
Although there is significant support in literature for the efficacy of an EMS device, its use has not yet become widespread. This maybe due to the poorly defined indications for its use. A number of papers have advocated their use in the following patient populations who usually have low fusion rates.
- Patients undergoing multilevel fusions
- Patients undergoing revision surgery
- Patients who smoke
The use of EMS devices in the first two groups produces a clinically and statistically significant increase in fusion rates.28 Although EMS showed an increased fusion rate in smokers, the numbers did not achieve statistical significance. This may be due to the fact that no study has been designed to directly assess this question, and the only data available on this topic are from other studies where smokers are a small subgroup.12,15,27,28
Of the three modalities of EMS, DCES is supported by the most scientific evidence and clinical trials. However, DCES is invasive and adds time and effort to the surgery. The efficacy of PEMF and CCES is less clearly proven but, as external devices, they can be easily applied. However, this also constitutes a drawback as the efficacy of these two methods becomes dependent on patients' compliance.
Studies comparing the efficacy and establishing the dosing of the three modalities must be undertaken. Although one study tried to compared DCES to PEMF as adjuncts to spinal fusion, it found that both methods have no statistically significant effect on rate of fusion.43One problem marring the study on EMS in spinal fusion is the lack of consensus over validated criteria defining fusion and pseudarthrosis, and the different endpoints used by various trials. Many methods for assessing fusion and pseudarthrosis have been used; however, they have all shown error and high false-negative results.44 The most accurate way of assessing pseudarthrosis is re-exploration15,45; however, this is not possible in clinical trials.
Most studies report fusion rates significantly lower than those seen in practice by most clinicians. This could be due to the fact that studies report lower figures than average or that patients are scrutinized for fusion more meticulously in a study compared to the average clinic.
With a price ranging from $3,000 to $7,000, EMS devices are associated with a significant cost that one might argue is unwarranted. However, one study found that after a 2-year follow-up, patients fitted with DCES devices showed significant long-term cost savings compared to those without DCES devices.46 The study went on to suggest that this may be due to higher fusion rates that reduce reoperation rates and overall morbidity associated with spinal fusion surgery.47
Direct current electrical stimulation has been reviewed by several insurance companies and was found to improve outcomes among patients at high risk for pseudarthrosis.47
Conclusion
Electromagnetic stimulation is a promising modality to assist in spinal fusion, and it is currently used by some spinal surgeons. There seems to be enough clinical evidence to advocate wider use. However, until scientific literature reaches a consensus about all EMS modality dosing and criteria for use, it will continue to be critically scrutinized by many surgeons.
References
- France JC, Norman TL, Santrock RD, McGrath B, Simon BJ. The efficacy of direct current stimulation for lumbar intertransverse process fusions in an animal model. Spine. 2001; 26:1002-1008.
- Bernard TN Jr. Repeat lumbar spine surgery. Factors influencing outcome. Spine. 1993; 18:2196-2200.
- Stauffer RN, Coventry MB. Posterolateral lumbar-spine fusion. Analysis of Mayo Clinic series. J Bone Joint Surg Am. 1972; 54:1195-1204.
- Friedenberg ZB, Harlow MC, Brighton CT. Healing of nonunion of the medial malleolus by means of direct current: A case report. J Trauma. 1971; 11:883-885.
- Dwyer AF, Wickham GG. Direct current stimulation in spinal fusion. Med J Aust. 1974; 1:73-75.
- Friedenberg ZB, Brighton CT. Bioelectric potentials in bone. J Bone Joint Surg Am. 1966; 48:915-923.
- Borgens RB. Endogenous ionic currents traverse intact and damaged bone. Science. 1984; 225:478-782.
- Otter MW, McLeod KJ, Rubin CT. Effects of electromagnetic fields in experimental fracture repair. Clin Orthop. 1998; 355(Suppl):S90-S104.
- Shamos MH, Lavine LS. Physical bases for bioelectric effects in mineralized tissues. Clin Orthop. 1964; 35:177-188.
- Bassett CA, Becker RO. Generation of electric potentials by bone in response to mechanical stress. Science. 1962; 137:1063-1064.
- Shamos MH, Lavine LS, Shamos MI. Piezoelectric effect in bone. Nature. 1963; 197:81.
- Oishi M, Onesti ST. Electrical bone graft stimulation for spinal fusion: A review. Neurosurgery. 2000; 47:1041-1055; Discussion 1055-1056.
- Friedenberg ZB, Andrews ET, Smolenski BI, Pearl BW, Brighton CT. Bone reaction to varying amounts of direct current. Surg Gynecol Obstet. 1970; 131:894-899.
- Friedenberg ZB, Zemsky LM, Pollis RP, Brighton CT. The response of non-traumatized bone to direct current. J Bone Joint Surg Am. 1974; 56:1023-1030.
- Kahanovitz N. Electrical stimulation of spinal fusion: A scientific and clinical update. Spine J. 2002; 2:145-150.
- Cundy PJ, Paterson DC. A ten-year review of treatment of delayed union and nonunion with an implanted bone growth stimulator. Clin Orthop. 1990; 259:216-222.
- Chou CK, McDougall JA, Chan KW. RF heating of implanted spinal fusion stimulator during magnetic resonance imaging. IEEE Trans Biomed Eng. 1997; 44:367-373.
- Shellock FG, Hatfield M, Simon BJ, et al. Implantable spinal fusion stimulator: assessment of MR safety and artifacts. J Magn Reson Imaging. 2000; 12:214-223.
- Becker RO. Stimulation of partial limb regeneration in rats. Nature. 1972; 235:109-111.
- Becker RO. The significance of electrically stimulated osteogenesis: More questions than answers. Clin Orthop. 1979; 141:266-274.
- Bassett CA, Pawluk RJ, Becker RO. Effects of electric currents on bone in vivo. Nature. 1964; 204:652-654.
- O'Connor BT. Effects of electric current on bone in vivo. Nature. 1969; 222:162-163.
- Luben RA. Effects of low-energy electromagnetic fields (pulsed and DC) on membrane signal transduction processes in biological systems. Health Phys. 1991; 61:15-28.
- Norton LA, Rodan GA, Bourret LA. Epiphyseal cartilage cAMP changes produced by electrical and mechanical perturbations. Clin Orthop. 1977; 124:59-68.
- Spadaro JA. Electrically stimulated bone growth in animals and man. Review of the literature. Clin Orthop. 1977; 122:325-332.
- Nerubay J, Marganit B, Bubis JJ, Tadmor A, Katznelson A. Stimulation of bone formation by electrical current on spinal fusion. Spine. 1986; 11:167-169.
- Kane WJ. Direct current electrical bone growth stimulation for spinal fusion. Spine. 1988; 13:363-365.
- Rogozinski A, Rogozinski C. Efficacy of implanted bone growth stimulation in instrumented lumbosacral spinal fusion. Spine. 1996; 21:2479-2483.
- Meril AJ. Direct current stimulation of allograft in anterior and posterior lumbar interbody fusions. Spine. 1994; 19:2393-2398.
- Kucharzyk DW. A controlled prospective outcome study of implantable electrical stimulation with spinal instrumentation in a high-risk spinal fusion population. Spine. 1999; 24:465-468; Discussion 469.
- Bassett CA, Pawluk RJ. Noninvasive methods for stimulating osteogenesis. J Biomed Mater Res. 1975; 9:371-374.
- Cain CD, Adey WR, Luben RA. Evidence that pulsed electromagnetic fields inhibit coupling of adenylate cyclase by parathyroid hormone in bone cells. J Bone Miner Res. 1987; 2:437-441.
- Ferrier J. Osteoclasts and osteoblasts migrate in opposite directions in response to a constant electrical field. J Cell Physiol. 1986; 129:283-288.
- Fitzsimmons RJ. Low-amplitude, low-frequency electric field-stimulated bone cell proliferation may in part be mediated by increased IGF-II release. J Cell Physiol. 1992; 150:84-89.
- Yen-Patton GP. Endothelial cell response to pulsed electromagnetic fields: Stimulation of growth rate and angiogenesis in vitro. J Cell Physiol. 1988; 134:37-46.
- Bodamyali T, Bhatt B, Hughes FJ, et al. Pulsed electromagnetic fields simultaneously induce osteogenesis and upregulate transcription of bone morphogenetic proteins 2 and 4 in rat osteoblasts in vitro. Biochem Biophys Res Commun. 1998; 250:458-461.
- Mooney V. A randomized double-blind prospective study of the efficacy of pulsed electromagnetic fields for interbody lumbar fusions. Spine. 1990; 15:708-712.
- Linovitz RJ. Fusion outcome. J Spinal Disord. 2000; 13:273.
- Zhuang H, Wang W, Seldes RM, Tahernia AD, Fan H, Brighton CT. Electrical stimulation induces the level of TGF-beta1 mRNA in osteoblastic cells by a mechanism involving calcium/calmodulin pathway. Biochem Biophys Res Commun. 1997; 237:225-229.
- Lorich DG, Brighton CT, Gupta R, et al. Biochemical pathway mediating the response of bone cells to capacitive coupling. Clin Orthop. 1998; 350:246-256.
- Brighton CT, Luessenhop CP, Pollack SR, Steinberg DR, Petrik ME, Kaplan FS. Treatment of castration-induced osteoporosis by a capacitively coupled electrical signal in rat vertebrae. J Bone Joint Surg Am. 1989; 71:228-236.
- Goodwin CB, Brighton CT, Guyer RD, Johnson JR, Light KI, Yuan HA. A double-blind study of capacitively coupled electrical stimulation as an adjunct to lumbar spinal fusions. Spine. 1999; 24:1349-1356; Discussion 1357.
- Jenis LG, An HS, Stein R, Young B. Prospective comparison of the effect of direct current electrical stimulation and pulsed electromagnetic fields on instrumented posterolateral lumbar arthrodesis. J Spinal Disord. 2000; 13:290-296.
- Blumenthal SL, Gill K. Can lumbar spine radiographs accurately determine fusion in postoperative patients? Correlation of routine radiographs with a second surgical look at lumbar fusions. Spine. 1993; 18:1186-1189.
- Guizzardi S, Di Silvestre M, Govoni P, Scandroglio R. Pulsed electromagnetic field stimulation on posterior spinal fusions: a histological study in rats. J Spinal Disord. 1994; 7:36-40.
- Hodges SD, Eck JC, Humphreys SC. Use of electrical bone stimulation in spinal fusion. J Am Acad Orthop Surg. 2003; 11:81-88.
- Kahanovitz N. Kahanovitz N. The role of implantable direct current stimulation in the critical pathway for lumbar spinal fusion. The Journal of Care Management. 1996; 2(6).